
Nitrogen determination is an important quality control test for pharmaceutical ingredients and other materials with specified nitrogen content. Accurate results depend on following the applicable analytical procedure, controlling digestion and titration conditions, and calculating results on the correct reporting basis.
NJ Labs provides USP <461> testing services for APIs, excipients, and raw materials, using the Kjeldahl procedure the chapter describes. When a specification allows combustion instead, we also run Dumas nitrogen analysis. We are an independently owned, FDA- and DEA-inspected CGMP laboratory and have been testing since 1936. We provide nitrogen determination testing in accordance with the agreed testing scope and applicable specification.
USP General Chapter <461> Nitrogen Determination describes the Kjeldahl procedure for nitrogen content determination in organic substances. It describes a macro procedure (Method I) and a semimicro procedure (Method II), and Method I includes a variation for samples that contain nitrates or nitrites. Individual monographs point to the chapter. The chapter gives the procedure; the monograph gives the sample preparation, the basis the result is reported on, and the acceptance limits. Povidone and heparin sodium are two familiar examples.
Monographs get revised, so check the current USP-NF for your material before you request testing. If you’re not sure which edition applies, send us the monograph or specification and we’ll work from it.
Kjeldahl nitrogen analysis converts the nitrogen in a sample to ammonia and then measures the ammonia. It runs in three stages:
Two practical points. A reagent blank runs alongside the samples, because trace nitrogen in the reagents shifts low-level results. And monograph limits are often written on a dried or anhydrous basis, so we can run moisture on the same lot, by Karl Fischer or loss on drying depending on the monograph, and report the corrected value.
The method also has limits. USP <461> notes that some alkaloids and other nitrogen-containing organic compounds don’t release all of their nitrogen on sulfuric acid digestion, so the Kjeldahl procedure doesn’t suit every material. It’s one reason we ask what the material is before we quote.
| Sample type | Why nitrogen is measured |
|---|---|
| Excipients with a nitrogen assay (povidone, for example) | The monograph uses nitrogen content as the assay, which reflects polymer composition. |
| Organic APIs and biologically derived materials (heparin sodium, for example) | The monograph sets a nitrogen specification the lot must meet. |
| Incoming raw materials | Verifies a lot against the nitrogen value on the supplier’s CoA or in your own specification. |
| Supplement and botanical ingredients | Total nitrogen is the starting point for a protein estimate once the right conversion factor is applied. |
Kjeldahl and Dumas are the two established nitrogen analysis methods, and NJ Labs runs both. Kjeldahl is the wet chemistry method described in USP <461>, so it’s the method we use when your monograph or specification references the chapter. The Dumas method burns the sample at high temperature and measures the nitrogen released. We run it on a Thermo Scientific FLASH 2000 Organic Elemental Analyzer. It’s faster, needs no digestion or distillation, and measures total nitrogen, including forms Kjeldahl digestion can miss. It’s often the practical choice for supplement, botanical, and food ingredients where the specification allows combustion or you need a protein estimate.
If your specification names USP <461>, tell us before you send samples. A Dumas result isn’t automatically interchangeable with a Kjeldahl result, and it’s better to settle which method applies before work starts.
Nitrogen Analyzer: https://njlabs.com/instrumentation/nitrogen-analyzer/
Nitrogen determination requires careful execution of the applicable procedure, appropriate analytical controls, and accurate calculations. NJ Labs offers wet chemistry testing and complementary analytical services to support pharmaceutical quality control and raw material evaluation.
Send us your material for USP <461> analytical testing, along with the monograph or specification and the reporting basis you need. We will confirm the test scope, sample quantity, and pricing. All project discussions are handled under confidentiality.
It is the compendial Kjeldahl procedure for measuring nitrogen in organic substances, described in USP General Chapter <461> and called for by individual monographs. The result is compared against the acceptance limits in the monograph.